Residual Field Indicator (Gauss Meter) – Full Description
A Residual Field Indicator, also known as a Gauss Meter, is a handheld measuring instrument used to check the amount of residual magnetism (remanence) present on a ferromagnetic component after Magnetic Particle Testing (MPI).
It ensures that the part has been demagnetized properly and meets acceptance standards.
✔ Purpose
A Residual Field Indicator is used to:
-
Measure remaining magnetic field after demagnetization
-
Ensure residual field is within acceptable limits (usually < 3 Gauss)
-
Identify areas where magnetism may cause:
✔ Interference with service performance
✔ Problems during welding, assembly, machining
✔ Attraction of metal debris in service
✔ What It Measures
-
Magnetic Flux Density in Gauss (G) or Tesla (T)
-
Residual magnetism left in the part
-
Polarity of magnetic field
-
Leakage fields on steel surfaces
A Residual Field Indicator is typically single-axis, meaning it measures field strength in one direction at a time.
✔ Construction & Components
-
Hall-Effect Probe
-
Sensitive to magnetic field
-
Usually transverse-type for surface measurements
-
Connected by shielded cable
-
-
Digital Display Unit
-
Provides numeric reading of Gauss/Tesla
-
Includes buttons for zeroing and unit selection
-
-
Zero Adjustment
-
Removes ambient magnetic influence before measurement
-
-
Power Source
-
Battery-operated for field use
-
✔ Typical Specifications
| Parameter | Typical Value |
|---|---|
| Measurement Range | 0 to ±20,000 Gauss |
| Resolution | 0.1 to 1 Gauss |
| Accuracy | ±1% to ±2% |
| Display | Digital LCD |
| Probe Type | Hall-effect, single-axis |
| Units | Gauss / Tesla |
| Operating Temp | 0–50°C |
✔ How It Is Used in MPI (Step-by-Step)
-
Zero the Meter
-
Hold probe away from any magnetic object
-
Press “ZERO” or adjust manually
-
-
Place Probe on Component Surface
-
Keep probe flat and stable
-
Avoid vibration and movement
-
-
Read Gauss Value
-
Positive (+) or Negative (−) indicates polarity
-
Value indicates field strength
-
-
Compare With Acceptance Criteria
-
Commonly required residual field: < 3 Gauss
(As per ASME, ASTM E1444, ISO 9934 guidelines)
-
-
If High Residual Magnetism Is Found
-
Repeat demagnetization
-
Recheck until acceptable
-
✔ Applications
-
Magnetic Particle Inspection (MPI)
-
Demagnetization verification
-
Checking magnetic leakage around welds
-
Evaluating residual field in bearings, shafts, gears
-
Ensuring safe assembly of precision parts
-
Determining if magnetic field may cause service problems
✔ Advantages
-
High accuracy
-
Portable and easy to use
-
Fast response
-
Essential for NDT (MPI) quality control
✔ Limitations
-
Affected by probe orientation
-
Must be zeroed frequently
-
Single-axis only (except advanced models)
-
Requires calibration annually
If you want, I can also provide:
📌 ALT image description of Residual Field Indicator
📌 Difference between Residual Field Indicator & Gauss Meter
📌 Calibration procedure
📌 Demagnetization acceptance criteria table
Tell me what you need!